Lifetime Hours
Lifetime Hours
A message can be defined that will be transmitted whenever the scanner's elapsed
time clock exceeds the set hours threshold. The message repeats one time every 30
minutes.
Hours Threshold
Definition:
A user can set a time in hours that will cause a message to be sent
whenever the scanner's run time exceeds this value.
The elapsed time is the time the scanner is powered up.
Usage:
Useful for setting up a preventive maintenance program.
Serial Cmd:
<K$
hours10,message>
Default:
2500
(x 10 = 25,000 hours)
Options:
0 to 65534
Note: Hours entered are 1/10 of the actual hours.
Message
Definition:
A user can define message up to 10-characters that will be displayed
whenever the hours threshold is reached or exceeded.
Serial Cmd:
<K$hours10,message>
Default:
SERVICE
Options:
Any 1 to 10 character ASCII string except NUL, <, or >.
10-10