Computer Interface - Hanna Instruments HI 8510 Instruction Manual

Panel - mounted orp - ph indicators and controllers
Hide thumbs Also See for HI 8510:
Table of Contents

Advertisement

installed without special connectors.
MAXIMUM 50 METERS
WET ENVIRONMENT
RELATIVE HUMIDITY UP TO 100%
®
AmpHel
electrodes have a micro-amplifier in
the electrode cap to boost the signal, drasti-
cally reducing susceptibility to noise and drift.
With all of the components sealed in the
electrode body, moisture up to 100% RH will
not effect the signal.
LONG DISTANCE, INDOOR/OUTDOOR IN-
STALLATIONS, ISOLATED OUTPUT FOR

COMPUTER INTERFACE

If your application has a distance greater
than 50 m (165'), it is necessary to install a
transmitter between the electrode and the
meter.
Hanna offers a full line of pH/ORP transmit-
ters with or without LCD displays.
MAXIMUM 300 METERS
3m
WET ENVIRONMENT
RELATIVE HUMIDITY UP TO 100%
AmpHel
®
is a registered Trademark of "Hanna Instruments"
57
pH-ORP METER
pH METER
WITH AN AmpHel
®
ELECTRODE
MEASUREMENTS CAN BE
TAKEN FROM DISTANCES UP TO
50 METERS (165 FEET)
pH-ORP METER
pH METER
WITH A pH TRANSMITTER AND
CONVENTIONAL ELECTRODE
MEASUREMENTS CAN BE
TAKEN FROM DISTANCES UP TO
300 METERS (1000 FEET)