Table A-2: Status Codes - Nellcor NPB-190 Service Manual

Pulse oximeter
Table of Contents

Advertisement

Appendix
A4.2 Patient Data and Operating Status
A-4
Time Tag
Time Tag
%SpO2
123456789
100
Time Tag does not represent a real-time clock. The number beneath the Time
Tag heading represents time, in seconds, since the unit was initialized at the
factory. This number will increase in size throughout the life of the monitor.
Patient Data
NPB-190
Version 1.0.0
Time Tag
%SpO2
123456795
100
Patient data are highlighted in the display above. Parameter values are displayed
directly beneath the heading for each parameter. In this example, the %SpO
100, and the pulse rate PR is 190 beats per minute. The "*" next to the 190
indicates that 190 beats per minute is outside of the alarm limits, indicated in the
top row, for pulse rate. If no data for a parameter is available three dashes (- - -)
will be displayed in the printout.
PA is an indication of pulse amplitude. The number can range from 0 to 254 and
will typically range around 45. There are no alarm parameters for this value. It
can be used for trending information. It is an indication of a change in pulse
volume, pulse strength, or circulation.
NPB-190
Version 1.0.0
Time Tag
%SpO2
123456795
100
The Status column indicates alarm conditions and operating status of the NPB-
190. A Pulse High alarm is indicated by the PH in this example. The status
column can have as many as four codes displayed in one line of data. A complete
listing of the status codes is listed in Table A-2.
Code
LB
AS
AO
SD
PS
LP
SH
SL
PH
PL
Note: A Sensor Disconnect will also cause three dashes (- - -) to be displayed
in the patient data section of the printout.
PR (bpm)
PA
Status
120
220
CRC XXXX
SpO2 Limit:
PR (bpm)
PA
Status
190*
220
CRC XXXX
SpO2 Limit:
PR (bpm)
PA
Status
190*
220

Table A-2: Status Codes

Meaning
Low Battery
Alarm Silence
Alarm Off
Sensor Disconnect
Pulse Search
Loss of Pulse
Sat High Limit Alarm
Sat Low Limit Alarm
Pulse rate High Limit Alarm
Pulse rate Low Limit Alarm
30-100%
PR Limit: 100-180 bpm
PH
30-100%
PR Limit: 100-180 bpm
PH