Error Codes - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Troubleshooting

Error Codes

When the N-600x detects an error condition, it may display "EEE"
followed by an error code.
Note:The "XXX" indicates the error code number may contain up to
three digits.
When an error code (other than the ones listed in Table 12) is
displayed, turn the instrument off and back on again. If the error code
reappears, record it and notify service personnel. Below Table 12 lists
the error codes and possible causes. When this occurs, the unit will
stop monitoring, remove all information from the screen and display
the message "EEE XXX," and sound a low priority alarm. Cycling
the power clears these errors.