Fuse Replacement; Figure 6-1: Fuse Removal - Nellcor NPB-190 Service Manual

Pulse oximeter
Table of Contents

Advertisement

Section 6: Disassembly Guide
6.3

FUSE REPLACEMENT

6-2
1.
Complete the procedure in paragraph 6.2.
2.
Disconnect the power cord from the back of the monitor.
3.
Remove the fuse drawer from the Power Entry Module by pressing down on
the tab in the center and pulling the drawer out as shown in Figure 6-1.
4.
Put new 0.5 amp fuses in the drawer and reinsert the drawer in the power
module.

Figure 6-1: Fuse Removal