6.8

POWER SUPPLY REMOVAL/INSTALLATION

6.
Position the ground line from the PEM so that it does not come into contact
with components on the Power Supply PCB.
7.
Complete procedure in paragraph 6.5.
Removal
1.
Complete the procedure described in paragraphs 6.2 and 6.4.
2.
Disconnect the leads from the battery.
3.
Follow the procedure in paragraph 6.7, steps 2 and 3.
4.
Use a 10mm wrench to disconnect the Power Supply ground lead from the
equipotential lug (Figure 6-6).
5.
Remove the seven screws shown in Figure 6-7.
6.
Lift the Power Supply out of the bottom case.

Figure 6-6: Power Supply Leads Connections

W1
W2
Equipotential
to
Brown to
Equipotential
"L" on
Lug
PEM
G
N
L
W4
W5
Red
Black
Section 6: Disassembly Guide
W3
Lug
Blue to
"N" on
PEM
6-7