Checking an actuator

Switch the electronics OFF on the remote.
Remove the controller cover, as described in chapter 8.8.3.
Take exact note of the positions of all cables and the sockets that they are connected to. Mark
the connectors and sockets or take a photograph with a digital camera.
Disconnect the motor plug (1) of the motor to be tested from the actuator module.
Check the electrical resistance at the
plug (1) of the actuator.
The plug can have a different shape than
show in the illustration.
If the resistance is close to infinity, the motor
is probably burnt out.
If the resistance is below 1 W, the motor has
a short-circuit.
The motor must be replaced in both cases.