Information Gauge Diagnostics - Invacare ProntoM41 Service Manual

With surestep
Table of Contents

Advertisement

SERVICE MANUAL
5.1.2

Information gauge diagnostics

Display
16
Description
1 Informtion gauge
All LEDs are off.
All LEDs are on.
Left RED LED is
flashing.
Left to Right "chase"
alternating with steady
display.
All LEDs are flashing
slowly.
Invacare
Definition
Power is off.
Power is on.
Battery charge is
low.
Joystick is in
programming, inhibit
and/or charging
mode.
Joystick has
detected Outof-
Neutral-at-Power-Up
mode.
®
- PRONTO M41
Comments
Fewer than three
LEDs on implies
reduced battery
charge.
The batteries
should be charged
as soon as
possible.
The steady LEDs
indicate the current
state of the battery
charge.
Release the
joystick back to
Neutral.