6.13 CHOOSING A LANGUAGE

The default language is English. If a change to another
language is desired, enter supervisory mode (See Section 6.1
Programming Options) and use the MODE button to scroll
down to:
To enter the menu, depress the ON/OFF button and use the
MODE button to select English, Spanish, French or German.
Depress ON/OFF to accept the selection and exit the menu.
6.14 SELECTING IMMEDIATE ALARM MODE/
There are two types of alarm modes to choose from:
immediate alarm mode and TWA alarm mode.
6. PROGRAMMING
TWA ALARM MODE
In the immediate alarm mode, two types of alarms are
available: immediate low and immediate high alarms. The
default values for these alarms are pre-set (See Appendix A
Specifications - Section (I) User Programmable High and
Low Alarm Set Points) at the factory but are user
programmable.
In the TWA alarm mode, three types of alarms are
available: TWA, STEL, and instantaneous alarms. TWA is
a time-weighted average over an 8-hour period. STEL is
the short-term exposure limit over a 15-minute period. The
instantaneous alarm represents an exposure level that
generates an alarm instantly. The values for these alarms
are pre-set (See Appendix A Specifications - Section (J)
Time Weighted Averages (Only if "TWA ON" is selected)
(Toxic sensors only)) at the factory and cannot be changed.
LANGUAGE SET