Honeywell MicroMAX Pro User & Maintenance Manual page 21

Multi-gas monitors
Table of Contents

Advertisement

MAN-MAXPRO
REV. B.0.
Instructions for using alarm check gas (part number Test 1A)
4. OPERATING INSTRUCTIONS
Turn monitor on in fresh air and allow it to go into normal
operating mode.
Break shipping tab off canister.
Attach blue balloon on to black disk on alarm check gas
hose. (See Figure 5)
Insert red end of hose into hole in canister nozzle.
Pinch hose (or place finger over the end of the hose) and
pulse trigger so that gas fills the balloon to its natural shape.
Do not over-inflate balloon.
Insert other end of hose into air intake.
All channels should be in alarm for a few seconds while gas
passes over sensors. Check display to confirm proper
operation of visual alarms for each gas. Confirm proper
operation of audible alarms.
Remove hose from monitor and allow alarm check gas to
be purged from monitor with fresh air.
Clear Peaks or, if "ALARM TYPE" is set to "TWA ON", clear
memory (Please see Section 4.7 Clearing Memory).
Gas/Bump test should be performed in a well-ventilated
area.
CAUTION
21