Carrier Gas Parameters
Parameter
CARRIER (He)
From 0.1 to 50.0 ml/min
Col.Flow
in steps of 0.1 ml/min;
Off
From 10 to 1000 kPa
Pressure
(from 2.5 to 250 kPa)
in step of 1 kPa;
Off
Constant Flow;
Flow Mode
Constant Pressure
Instruction Manual
CARRIER (He)
Pressure
(100)
Col.Flow
12.0 12.0
Flow mode
Const flow
Gas saver flow
Saver time
Vacuum comp.
Leak check
Column evaluation
Carrier Menu in
Constant Flow Mode
Table 3-1. Carrier Gas Parameters Menu
Range
This line is the title bar.
It is displayed when Constant Flow is selected in
Flow Mode. Enter a column flow for the analytical
column.
It is displayed when Constant Pressure is selected
in Flow Mode. Enter inlet pressure for the
analytical run.
Set the operating mode required.
Selecting Constant Flow, the column flow is kept
constant throughout the analysis. The pressure at the
column head will change with the column
temperature to maintain a consistent flow.
Selecting Constant Pressure, the pressure at the
column head is kept constant throughout the
analysis. During a temperature program, the column
flow decrease due to the increase of the carrier gas
viscosity.
CARRIER (He)
Pressure
Col.Flow
Flow mode Const Pres
100
Gas saver flow
1.00
Saver time
Off
Vacuum comp.
Leak check
Column Evaluation
Constant Pressure Mode
Description
Chapter 3
Gas Control
100 100
(10.0)
100
1.00
Off
Carrier Menu in
81