Section 8 Cartridge Preparation And Purification Protocol - BIO RAD Bio-Scale Instruction Manual

Cartridges 1 and 5 ml
Table of Contents

Advertisement

5.
Remove the supernatant, and filter through a
0.45 µM filter before applying to the cartridge.
Section 8
Cartridge Preparation and
Purification Protocol
To prepare a cartridge, remove the end plugs and
connect the cartridge to the chromatography system. The
cartridge is now ready for use. The following 1 ml
cartridge purification protocol is a general guideline
for first time users. Flow rates for 5 ml cartridges are
shown in parentheses. The kinetics of cleavage are
target protein specific and may require optimization
to maximize yield. See Figure 6 for a representative
purification profile using the following protocol. Refer
to the Profinity eXact System Manual for complete
details.
20
Purification Protocol
1.
Equilibrate the cartridge with 10 column volumes
(CV) of bind/wash buffer at 3 ml/min (10 ml/min).
2.
Load the sample lysate at 1 ml/min (5 ml/min).
3.
Wash the cartridge with 10 CV of bind/wash
buffer at 3 ml/min (10 ml/min).
4.
Equilibrate cartridge with 2 CV of elution buffer
at 3 ml/min (10 ml/min)*.
5.
Stop flow and incubate for 30 minutes
6.
Elute purified protein with 5 CV of elution buffer
at 3 ml/min (10 ml/min).
* Alternative elution protocol: Pump 5 CV of elution buffer at
0.1 ml/min (0.5 ml/min) at room temperature.
** Increased incubation time, 2–4 hr at 25°C or 12–24 hr at 4°C,
may be necessary for complete cleavage of some proteins.
**
.