Download Print this page

Application Examples - Kipp & Zonen LOGBOX SD Instruction Manual

Advertisement

18. Application examples

Example 1.
Connection of relative humidity and temperature sensors.
Example 2.
Memory capacity calculation
Let's assume we measure and store data from channels AIN1, AIN2 and DIN1 every 5
minutes. Now calculate how many bytes will take one record:
Date – 4 bytes
Time – 4 bytes
AIN1- 4 bytes
AIN2 – 4 bytes
DIN1 – 4 bytes
Together it is 20 bytes for one record. Internal memory capacity is 128kB (about
131000 bytes) and therefore we can store
131000/20=6550 records
Every 5 minutes record will make 12 records per hour. Nr. of recorded hours:
6550/12=545 hours and it is about 22 days (more than 3 weeks).