Digital inputs configuration commands
BE
Clear all settings for bipolar inputs
B[X]C1 (X represent digital input. X [1...4])
Enable digital input X for counter function
B[X]C0 (X represent digital input. X [1...4])
Disable digital input X for counter function
B[X]F1 (X represent digital input. X [1...4])
Enable digital input X for frequency function
B[X]F0 (X represent digital input. X [1...4])
Disable digital input X for frequency function
B[X]P[Y]
(X represents digital input, Y represents nr. of polynomial. X [1...4],
Y[0...8])
Set polynomial coefficient of P[Y] to digital input B[X]
If Y=0, no polynomial is used
Polynomial coefficients configuration commands
PE
Clear all settings for polynomial coefficient
P[X][Y]
(X represents indexed nr. of polynomial, Y represents coefficient.
Y[any real number])
Set up coefficients of polynomial
P40
=0
OK
>P41
P41
=10000
OK
>P42
P42
=125E-7
OK
>P43
P43
-321.245
OK
P?
Returns all coefficients of all polynomials
P?
1: 0 2.25 0.0009813 0
2: Not set.
3: Not set.
4: 0 10000 125E-7 –321.245