This configuration allows you to connect up to 5 sensors on the analog input which
have a output up to 2,5Vdc.
-AIN1 & AIN6
-AIN2 & AIN6
-AIN3 & AIN6
-AIN4 & AIN6
-AIN5 & AIN6
NOTE: single ended 24 bits measurements require AIN(6) and not the GND as
return line
And it's possible to connect up to two sensors on the analog input which have an
output up to 3V.
-AIN7 & GND
-AIN8 & GND
Please note that AIN7 and AIN8 can only be used to connect sensors which have a
slightly higher voltage output up to 3Vdc (Like our UVS sensors)
An example how to configure the software. (which will be explained in chapter 14)