Sartorius GPC65-CW Operating Instructions Manual page 71

Pipette calibration balance
Table of Contents

Advertisement

Display
err 30
print fct. blocked
ref.wt. too light
upd. not possible
no num. value
xxxxx too low
xxxxx too high
too many characters
x = 1 :
err 10
x = 2 :
x = 3 :
x = 4 :
All segments displayed continuously
err 340
No WP
blocked
The special code
remains displayed
J
The weight readout changes constantly
The weight readout is obviously wrong
If any other errors occur, contact your local Sartorius Service Center.
Cause
Interface port for printer output is blocked
Error in storing reference weight
(with the counting or weighing-in-
percent application)
Reference sample updating not possible
(with the counting application)
Input wrong
(for any application program), e.g.,
alphabetic input not allowed
Input text too long
Key is stuck
Key pressed when turning on
the balance:
c, q, w, v
0, 3, 4, 9
2, 5, 6, p, t – right
1, 7, 8, a, t – left
Either M was pressed when you turned
on the balance, or this key is stuck
Operating parameter (EEPROM) is wrong
Weighing cell (platform) defective
Function blocked
None of the keys has been pressed
since the balance was turned on
Unstable ambient conditions
Too much vibration, or the balance
is exposed to a draft
A foreign object is caught between
the pan and the balance housing
The balance has not been calibrated/adjusted Calibrate/adjust the balance
The balance was not tared before weighing
The balance is not level
The dust cover is caught under the
weighing pan
Solution
Contact your local Sartorius Service Center
Weight too light or there
is no sample on the balance
See "Counting" in "Operation
the Balance" for reference updating criteria
Follow the instructions for the
application programs
Allowable text lengths, incl. decimal point:
– S ID and L ID: 20 characters max.
– W ID: 14 characters max. for weights
Contact your local Sartorius Service Center
Release key
Contact your local Sartorius Service Center
Contact your local Sartorius Service Center
None
Press a key
Set up the balance in another area
Change Setup configurations
to adapt the balance to the ambient conditions
Remove the foreign object
Tare before weighing
Level the balance
See "Replacing the Dust Cover" in the chapter
entitled "Care and Maintenance"