Moving The Patient - Invacare GET-U-UP LIFT Maintenance Manual

Invacare wheelchair/scooter lift owner's operator and maintenance manual
Table of Contents

Advertisement

SECTION 4—LIFTING
NOTE: The lower center of gravity provides stability making the patient feel more secure and the lift easier to move.
NOTE: The lift arms will stay in position until lowered.
Handgrips
Steering
Handle
Mast
Handle
Rear
Casters

Moving the Patient

DO NOT move a person suspended in a sling any distance. The Invacare lift is NOT a transport
device. It is intended to transfer a seated individual from one resting surface to another (such as
a bed to a wheelchair). Otherwise, injury or damage may occur.
The legs of the stand up lift MUST be in the maximum open position for optimum stability and
safety. If the patient is in a sling and it becomes necessary to move through a narrow passage,
close the legs of the stand up lift only as long as it takes to move through the passage. When the
stand up lift is through the passage, return the legs to the maximum open position.
DO NOT roll caster base over objects such as carpet, raised carpet bindings, door frames, or any
uneven surfaces or obstacles that would create an imbalance of the lift during transfer of a
patient suspended in the lift sling. This could cause the lift to tip over. Use the mast handle at all
times to push or pull the lift.
NOTE: For this procedure, refer to Detail "C" of FIGURE 4.1.
1.
Ensure the legs of the stand up lift are in the maximum open position. If not, open legs to the maximum open position.
2.
Move the stand up lift away from the surface using the steering handle.
3.
Slowly move the patient to the desired surface.
Get-U-Up
Lift
DETAIL "A" - LIFT
FIGURE 4.1 Lifting the Patient
DETAIL "B" - WHEEL LOCK
Wheel Lock
DETAIL "C" - LIFTING THE PATIENT
Sling
Knee Pad
Footplate
WARNING