Temperature Units - Omega OM-DAQLINK-TEMP User Manual

Table of Contents

Advertisement

Online Mode: For working with logger connected to PC. Data will be transmitted in real-
time i.e. at the same time as the sampling rate.
Optimized Memory Mode: For working with the logger as a standalone device, not
connected to the PC. In this mode, the memory is more efficiently managed as the data
does not need to be transmitted together with the sampling rate. Use the Transmissions
interval parameter in the Setup window to see the rate at which the data will be displayed
on the logger or in the software.
Note: If used when connected to the PC, the data will not be transmitted in real-time.
Cyclic Mode
If this checkbox is not selected, the logger will stop recording data when the memory capacity is
filled.
Enabling Cyclic mode will result in the oldest samples in the memory being overwritten by new
samples once the memory is full. This allows continuous data logging.
Averaging Points
You can choose to have online averaging of up to 10 points around the real value in order to
smooth the data readings, if they are exhibiting interefence. The recommended averaging is 4
samples.
Note: If configuring Alarm duration with averaging points selected, the alarm will be counted
from the time of the last sampling point used to calculate the average.

Temperature Units

Toggle between °C and °F. This applies to the readings displayed on the logger LCD as well as
in the software.
Anti Tamper Mode
To prevent tampering with the OM-DAQLINK logger, selecting this option will disable the
following features on the logger menu:
Pause
Reset
Stop and turn off unit
Active Sensors
Depending on the logger model, you can activate up to two internal sensors and four external
sensors. Simply select the inputs you want to use and then select the sensor types from the
adjacent drop-down menu. You don't have to select the external sensor input in any order. You
may select only Input 4 if you desire.
62

Advertisement

Table of Contents
loading

This manual is also suitable for:

Om-daqlink-temprh

Table of Contents