Summary of Contents for Omega OM-DLAC

  • Page 1 AC Current Data Logger...
  • Page 2: Om-Dlac Data Logger Instructions

    This interval can be set from 5 seconds to 60 minutes. A user message of up to 20 characters can also be stored in the OM-DLAC. This message can be used to identify a customer’s location or identify a particular logger for tracking pur- poses.
  • Page 3 Refer to the owner’s manual for your computer to find instructions on setting the PC time if it is not correct. 8. When all of the settings for the OM-DLAC are to your satisfaction, click on OK to store the setting in the OM-DLAC and to close the window.
  • Page 4: T O Start A Recording

    AC Current Data Logger 10. You can now connect a different OM-DLAC or exit the program. To exit the program, click on the Exit button on the tool bar. TART A ECORDING To begin a new recording, slide the access door closed. The LED indicator will begin to flash for 10 seconds.
  • Page 5: Connecting The Om-Dlac

    Plug each lead from the clamp on probe firmly into the OM-DLAC. Make sure that there is no strain that could pull the leads out of the OM-DLAC socket. Place the clamp on probe around the wire that the current is being measured. Make sure that the probe will not interfere with any moving pieces of equipment and will not pull loose the wire being measured.
  • Page 6: Viewing A Recording As A Graph

    AC Current Data Logger been using the text mode, then the recording will be displayed in the text mode. If you have been using the graph mode, then the recording will be displayed as an auto scaled graph. IEWING A ECORDING AS A RAPH When information is re-...
  • Page 7: View A Recording As Text

    AC Current Data Logger To auto-scale the graph, click on the Auto Scale button. The graph will automatically be readjusted to fill the vertical axis of the screen. The graph on the screen may be printed at any time by clicking on the Print button.
  • Page 8: Data Logger Calibration

    New Password section. Click the OK button when finished. If this is the first time you are calibrating the OM-DLAC, no pass- word is needed. You can click the OK button or press the Enter key to continue.
  • Page 9 The LED will stop flashing and the OM-DLAC will not allow a new record- ing to be started until the battery has been replaced. Recorded data can be re- trieved regardless of the condition of the battery.
  • Page 10 AC Current Data Logger OM-DLAC S PECIFICATIONS Ambient Operating Temperature +10°F to +150°F (-12°C to +65°C) Range Storage Temperature -40°F to +170°F (-40°C to +76°C) Time Accuracy +/- 100ppm @ 75°F Current Range 0 to 300 Amps AC using supplied...
  • Page 11: Om-Dp3 Data Plotter

    AC Current Data Logger OM-DP3 D LOTTER The OM-DP3 is an optional accessory instrument that allows the user to retrieve data and program data loggers in the field without the need of a computer. Using the OM-DP3 you can plot recorded information from a data logger or change the settings of a data logger prior to making a different data recording.