Omega CA450 User Manual page 68

Process multimeter
Table of Contents

Advertisement

Outputs setup data
OS
Command = OS<CRLF> -> Return =
Parameters
a (measurement) = ON or OFF
b (measurement function) = DCV, OHM, FREQ, ACV, DCmV,
c (measurement range) = 1000V, 600V, 60V, 6V for
d (output) = ON or OFF
e (output function) = DCmA
f (output range) = 20 mA for DCmA
g (output value)
h (output for 24 V loop measurement) = ON or OFF
i (backlight) = ON or OFF
j (charging) = OFF (always OFF)
Measure a<CRLF>
Function b<CRLF>
Range c<CRLF>
Source d<CRLF>
Function e<CRLF>
Range f<CRLF>
Data g<CRLF>
24V Output h<CRLF>
Light i<CRLF>
Charge j<CRLF>
ACmV, DIODE, CONTINUITY,
DCmA, LOOP POWER
DCV/FREQ
600 mV for DCmV
600 mV for ACmV
60 MOHM, 6MOHM, 600kOHM,
60kOHM, 6kOHM, 600OHM for OHM
2 V for diode test
600OHM for continuity check
100 mA or 30 mA for DCmA
30 mA for loop power