Sets or queries the MIN/MAX/AVG function
MI
Command = MIm<CRLF> -> Answer = MIm<CRLF>
Command = MI?<CRLF> -> Return = MIm<CRLF>
Parameters
m = 0: MIN/MAX/AVG recording is disabled (default)
If the CA450 is in an output mode, ERR13 is returned.
Sets or queries the span setting for the 100 mA range for
MP
DCmA measurement
Command = MPm<CRLF> -> Answer = MPm<CRLF>
Command = MP?<CRLF> -> Return = MPm<CRLF>
Parameters
m = 0: 0 to 100 mA (default)
If the CA450 is in an output mode, ERR13 is returned.
The changed setting is valid until it is changed again.
1: MIN/MAX/AVG recording starts
1: 10 to 50 mA
2: 0 to 50 mA
6
Normal Cal.
Î
Normal Cal.
Î