Omega CA450 User Manual page 37

Process multimeter
Table of Contents

Advertisement

Step
(Subdisplay)
-25%
0%
25%
50%
75%
100%
125%
131.2%
For the method for setting the current span, see section .., "Current Span Setting."
Span Check Mode
You can switch the output current between 20 mA and  mA (or 20 mA and 0 mA) using
just the %STEP keys.
When you hold down %STEP▲ or %STEP▼ for 1 second or longer, "SPAN" appears, and
the CA0 enters into span check mode.
In this mode, the output switches to 100% when you press %STEP▲ and to 0% when you
press %STEP▼.
Setting Key
%STEP
The COARSE (0. mA increase and decrease) and FINE (0.00 mA increase and decrease)
keys can also be used.
To exit span check mode, hold down %STEP▲ or %STEP▼ for 1 second or longer again.
("SPAN" disappears.)
Tip
• If the set current is 0. mA or more and the terminals are open, "-----" appears in the
main display, and "----" appears in the subdisplay.
Output Value
 to 20 mA
0.000 mA
.000 mA
8.000 mA
2.000 mA
6.000 mA
20.000 mA
2.000 mA
When the range is 4 to 20 mA
Output Setting
20.000 mA (100%)
 to 20 mA
4.000 mA (0%)

0 to 20 mA
0.000 mA
.000 mA
0.000 mA
.000 mA
20.000 mA
2.000 mA
0 to 20 mA
0.000 mA (0%)