Omega HHM23 User Manual

Omega HHM23 User Manual

Digital multimeter

Advertisement

OMEGA

Advertisement

Table of Contents
loading

Summary of Contents for Omega HHM23

  • Page 1 OMEGA...
  • Page 2 European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Safety Information

    SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter: 1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly. 2.
  • Page 4: Specifications

    SPECIFICATIONS Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999. Polarity: Automatic, positive implied, negative polarity indication. Overrange: (OL) or (-OL) is displayed. Zero: Automatic. Low battery indication: the " " is displayed when the battery voltage drops below the operating level.
  • Page 5 DC VOLTS Ranges: 200mV,2V,20V,200V,600V Resolution: 100µV Accuracy: ±(0.35%rdg + 1dgt) Input impedance: 10MW Overload protection: 500VDC or AC rms on 200mV range 600VDC or AC rms on all other ranges AC VOLTS (50Hz - 500Hz) Ranges: 200mV,2V,20V,200V,600V Resolution: 100µV Accuracy: ±(1.0%rdg + 4dgts) on 200mV to 200V ranges ±(2.0%rdg + 4dgts) on 600V range Input impedance: 10MW Overload protection: 500VDC or AC rms on 200mV range...
  • Page 6 AC CURRENT (50Hz - 500Hz) Ranges: 200µA,2mA,20mA,200mA,10A Accuracy: ±(1.2%rdg + 4dgts) on µA,mA ranges ±(3.5%rdg + 4dgts) on 10A range Input protection: 0.5A / 250V fast blow ceramic fuse 10A / 600V fast blow ceramic fuse RESISTANCE Ranges: 200W ,2KW ,20KW ,200KW ,2MW ,20MW Accuracy: ±(0.4%rdg + 4dgts) on 200W range ±(0.4%rdg + 1dgt) on 2KW to 2MW ranges ±(3.0%rdg + 4dgts) on 20MW range...
  • Page 7 CAPACITANCE Ranges: 2nF,20nF,200nF,2µF,20µF Accuracy: ±(2.0%rdg + 10dgts) Test frequency: 2.5Hz FREQUENCY (Autoranging) Ranges: 2KHz,20KHz,200KHz,2000KHz,15MHz Accuracy: ±(0.1%rdg + 1dgt) Sensitivity: 1.0Vrms min Overload protection: 500VDC or AC rms Effect Reading: 10 - 1999 LOGIC TEST Threshold: Logic Hi (2.8 ± 0.8V) Logic Lo (0.8 ±...
  • Page 8: Operation

    OPERATION Before taking any measurements, read the Safety Information Section. Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements.
  • Page 9 Current Measurements 1. Set the Function/Range switch to the desired current range and press the "AC/DC" switch toggle between to the desired current type. 2. For current measurements less than 200mA, connect the red test lead to the µAmA jack and the black test lead to the COM jack. 3.
  • Page 10 Diode Tests 1. Connect the red test lead to the "VW " jack and the black test lead to the "COM" jack. 2. Set the Function/Range switch to the " " position. 3. Turn off power to the circuit under test. 4.
  • Page 11 Transistor Gain Measurements 1. Set the Function/Range switch to the desired hFE range (PNP or NPN type transistor). 2. Never apply an external voltage to the hFE sockets. Damage to the meter may result. 3. Plug the transistor directly into the hFE socket. The sockets are labeled E, B and C for emitter, base, and collector.
  • Page 12: Maintenance

    MAINTENANCE WARNING Remove test leads before changing battery or fuse or performing any servicing. Battery Replacement Power is supplied by a 9 volt "transistor" battery. (NEDA 1604 IEC 6F22). The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery case.
  • Page 13 CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any...
  • Page 14: Data Acquisition

    Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE DATA ACQUISITION PRESSURE/STRAIN AND FORCE HEATERS FLOW/LEVEL ENVIRONMENTAL MONITORING AND CONTROL pH/CONDUCTIVITY M-2852/0799...

This manual is also suitable for:

Hhm24Hhm19c