23.10 VSI Remote Battery Gauge Diagnostics

The Flash code from the VSI battery gauge is shown below. The number of flashes is indicated by
the number of LED's illuminated
Display Description
All LED's off
All LED's on steady
7
Joystick Trip / Fault. Check for faulty bus cable or joystick
7
Joystick Trip / Fault. Check for faulty bus cable or joystick
7
Communications Error. Possibly Bus cable fault
7
Joystick Trip / Fault. Check for faulty bus cable or joystick
7
Joystick Trip / Fault. Check for faulty bus cable or joystick
7
Joystick Trip / Fault. Check for faulty bus cable or joystick
7
Communications Error. Possibly Bus cable fault
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Communications Error. Possibly Bus cable fault
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Omni + trip error. The Omni + unit is faulty. Replace.
Sip / Puff
Omni + Out of Calibration. The Sip / Puff input of the unit
needs to be calibrated.
7
Omni + is not receiving a valid signal for user input device.
7
Omni + is not receiving a valid signal for user input device.
7
Omni + is not receiving a valid signal for user input device.
7
Omni + is not receiving a valid signal for user input device.
7
Omni + is not receiving a valid signal for user input device.
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Communications Error. Possibly Bus cable fault
7
Omni + is not receiving a valid signal for user input device.
7
Omni + is not receiving a valid signal for user input device.
7
Omni + Out of Calibration. The Analogue input of the unit
needs to be calibrated.
7
Omni + trip error. The Omni + unit is faulty. Replace.
7
Dual Communications Trip. There is an error between the
joystick and dual control unit. Check cables.
7
Dual Communications Trip. There is an error between the
joystick and dual control unit. Check cables.
User
Joystick Displaced at power up. Allow joystick to re-center
then switch off and on again to clear.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
User
Dual Joystick Displaced at power up. Allow joystick to re-
center then switch off and on again to clear.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
7
Dual Attendant Module Trip. The unit is defective.
8
Possible power module Trip.
This Means That .....
Power is off
Power is ON
SERVICE MANUAL