Invacare
- Dragon Vertic
22.3
ACS2 Control System
Description:
1 ACS Remote
3 ACS Power Module
5 Bus Socket
Requirements:
· 7 mm open ended spanner
· 8 mm open ended spanner
· No 2 cross point screwdriver
· Flat blade screwdriver
22.3.1 Inspection
· Check for any physical damage to the outer case of Remote (1).
· Access the power module (3) and actuator module (4) as described in the batteries section of
this user guide and inspect for physical damage.
· Inspect the rubber gaiter around the joystick knob (2) for any cracks, splits or perishing. This
would enable water or moisture to access the unit. This may cause damage to the controller.
· Check the keypad is in good condition with no small tears or cracks.
· Check the bus cable (5) and connectors from the joystick remote (1) are free from damage.
2 Joystick Knob
4 Actuator Module
6 Keypad
SERVICE MANUAL
49