Pilot Plus Control System; Inspection; Disassembly; Remote - Invacare Spectra Blitz Service Instructions Manual

Hide thumbs Also See for Spectra Blitz:
Table of Contents

Advertisement

SERVICE MANUAL
22.4

Pilot Plus Control System

1
Description:
1 Pilot + Remote
2 Joystick Knob
3 Pilot + Power Module
Requirements:
· 7 mm open ended spanner
· 8 mm open ended spanner
· No 2 cross point screwdriver
· Flat blade screwdriver
· Terminal screwdriver

22.4.1 Inspection

· Check for any physical damage to the outer case of remote (1).
· Access the power module (3) and ALM module (4) as described in the batteries section of this
user guide and inspect for physical damage.
· Inspect the rubber gaiter around the joystick knob (2) for any cracks, splits or perishing. This
would enable water or moisture to access the unit. This may cause damage to the controller.
· Check the keypad is in good condition with no small tears or cracks.
· Check the bus cable (5) and connectors from the joystick remote (1) are free from damage.

22.4.2 Disassembly

Caution:
· Before any work is carried out on the control system, ensure the power module (3) is
unplugged from the batteries, and the batteries are disconnected.

22.4.2.1 Remote

· Unplug the Pilot Plus remote (1) from the power module (3) removing bus cable (5).
52
2
4 ALM Module
5 Bus Cable
Invacare
3
5
®
- Spectra Blitz
5

Advertisement

Table of Contents
loading