Floor Levelness

Environmental Effects

Laser Specifications

Power Requirements

January 17, 2007
2
G
/Hz
Freq. Range
-5
6 x 10
50-200 Hz
-6
5 x 10
200-650 Hz
-6
1 x 10
650-1000 Hz
The floor must be level within 1°.
• Heat Dissipation: total heat dissipation for the Imager is 4,800 BTU/hour
• Acoustical Noise:
– Less than or equal to 65 dB-A at 1 m during operation
– Less than or equal to 50 dB-A at 1 m during idle or standby mode
Item
Specification
Wavelength
810 nm
Power
120 mW
Radiation class 3B
Type
Diode
Four power cord sets are provided with this equipment:
• power cord with plug for use in North America
• power cord with plug for use in China
• power cord with plug for use in the United Kingdom
• power cord with plug for use in Europe
All other countries must use an Agency-approved power cord with plug type
suitable for the country of use, or contact an authorized Kodak dealer.
One of the following single-phase, 15 A power sources, with grounding, must
8F927
Specifications
Direction
Vertical or horizontal
Vertical or horizontal
Vertical or horizontal
6-3