Filter Maintenance

Cleaning the
Electronics Filter

Changing the PMM

January 17, 2007
There are two filters in the Laser Imager that must be attended to during the
preventive maintenance process.
The electronics filter should be cleaned every 6 months.
Air
intake
door
Electronics
Open the air intake (bottom) door.
1.
Use a vacuum cleaner to vacuum the filter.
2.
Close the air intake door.
3.
The Preventive Maintenance Module (PMM) must be changed every 15,000
films.
On the touch screen, touch:
1.
• the Unlock icon
• Unlock Film Supply
• OK
8F2927
Using and Maintaining the Imager
filter
2-11