Warnings Provided in the Instruction Manual
Injury from Being Pinched
• The specimen stage's weight is approximately 5 kg. It might cause injury when
operating it grasping the parts other than the handle. When changing specimens,
securely grasp the handle, and use caution so that your hands will not be pinched.
• The electron gun weighs is 2.5 kg. It might cause finger injury when opening and
shutting. Be sure not to pinch fingers when opening and shutting.
• The finger might get injured in the opening and shutting part of the top cover. Be sure
not to pinch fingers when opening and shutting.
Burns from Touching the Filament Assembly and Peripheral
Components
• After the system has been in use, the Wehnelt and the anode around the filament are
heated about 90℃. Touching any of them can cause burns. Before undertaking any
filament replacement work, turn off the main unit and wait at least 30 minutes to allow it
to cool down.
For the Safe Handling of the System
(continued)
CAUTION
SAFETY - 6
(Chapter 4. Operation)
(Chapter 5. Maintenance)
(Chapter 5. Maintenance)
(Chapter 5. Maintenance)