Notes:
*Exterior Overshield is connected to the shells of the connectors on both end s.
**3M p art 14X26-SZLB-XXX-0LC is a com p lete cable assem bly, inclu d ing connectors.
Unu sed p airs shou ld be term inated in 100 ohm s at both end s of the cable.
Inner shield is connected to signal grou nd insid e cam era
Table 9: Teledyne DALSA Camera Control Configuration
Signal
CC1
CC2
CC3
CC4
See Ap p end ix B for the com p lete Teled yne DALSA Cam era Link configu ration table, and refer to the
Know led ge Center
on Teled yne DALSA's Web site, for the official Cam era Link d ocu m ents.

Input Signals, Camera Link

The cam era accep ts control inp u ts throu gh the Cam era Link MDR26F connector.
The camera ships in internal sync, internal programmed integration (exposure mode 7) TDI Mode.
EXSYNC (Triggers Frame Readout)
Fram e rate can be set internally u sing the serial interface. The external cont rol signal EXSYN C is op tional
and enabled throu gh the serial interface. This cam era u ses the falling edge of EXSYN C to trigger p ixel
read ou t.
Direction Control
Control the CCD shift d irection throu gh the serial interface. Use the softw are com m and
w hether the d irection control is set via softw are control or via the Cam era Link control signal on CC3.
03-032-20117-01
3
CC3+
15
CC4+
2
CC4-
13
inner shield
26
inner shield
Configuration
EXSYN C
PRIN
Direction
Spare