ERROR MESSAGES

Low Battery
This message indicates that the battery should be
replaced soon. If the battery voltage drops too low,
continued use with a low battery can affect weighing
accuracy. Remove the old battery and replace with a
new battery for operation.
Over Load
The load is over the scale capacity. Reduce the load
and try again. If the trouble still exists, please call
your scale service representative.
Over Zero
The load is over the limit the scale can zero when
powering on. Reduce the load and try again. If the
trouble still exists, please call your scale service
representative.
Under Zero
The load is under the limit when powering on. Power
OFF scale and try again. If the trouble still exists, call
your scale service representative.