How To Convert Hp 339X Integrator Workfiles From 5890A To Series Ii Mode - HP 5890 Series II Plus Reference Manual

Hewlett-packard reference manual
Table of Contents

Advertisement

Signal Output
File compatibility with data handling devices
How to convert HP 339X Integrator workfiles from 5890A to
SERIES II mode:
1. Turn the GC off.
2. Follow the previous instructions to set the GC for 5890A mode (use
proper grounding).
3. Download the workfile from the integrator.
4. Turn GC power off.
5. Remove the P15 jumper. (Now the GC is in SERIES II mode.)
6. Turn GC power on.
7. Add SERIES II setpoints (time table, etc).
8. Store the workfile at the integrator.
Note: Using
workfiles can be converted with each jumper change sequence.
on the HP 5890 SERIES II, up to 4 integrator
STORE