HP 5890 Series II Plus Reference Manual page 230

Hewlett-packard reference manual
Table of Contents

Advertisement

Test Sample Chromatograms
Test sample chromatograms
Figure 10-6.
Detector Type
Temp 250 DEGREES C
Inlet Type SPLIT ONLY OR
SPLIT/SPLITLESS Temp
Operating Mode SPLIT(PURGE ON)
Purge Time On
Purge Time Off
Oven Temp Programmed (1 ramp)
Init Temp
Init Time
Ramp
Rate
Fin Temp
Fin Time
Range
7
COLUMN:
Part No.
19095Z•121
Dimensions
Sta Phase
Methyl Silicone
START
FID-Split Mode Capillary Inlet
230
HP 5890 Test Sample Operating Conditions
FIDw/MUG
200 DEG C
0
min
0
min
100
DEGREES C
0
min
20
180
1
¿
530
; 10 M
2.21 2.35
2.81
ST
FLOW RATES
Carrier (He)
15 +/• 1
Hydrogen
30 +/• 1
Air
400 +/• 20
Makeup (N2)
20 +/• 1
Split Vent
200 +/• 20
Septum Purge
5 +/• 1
SAMPLE:
Type
Capillary Sample
¿
Inj Volume
1
Part No. 18740•60900
Composition: Soln of 0.1%(w/w)
n-nonane,n-hexadecane and 0.1%
(w/w)each of 1-octanol,nonanol,
n-undecane,naphthalene,2-deca-
none,n-dadecane,n-tridecane,
1-pentadecane in tetradecane.
2.64
3.67
ml/min
ml/min
ml/min
ml/min
ml/min
ml/min
l
.40
C9
C16