Signal Output
INET configuration
Figure 5•1 1 shows resulting displays.
Figure 5-11
INET Signal Definition Displays
From the displays, the following may be noted:
HP 5890 signal channels are designated SIG 1 or SIG 2.
ON indicates the given signal channel is considered active by the
controller; data from this signal channel is transmitted to other
devices on the INET loop.
Similarly, OFF indicates the channel is considered inactive; no data
from this signal channel is transmitted to other devices on the loop.
It is important to note that ON or OFF in this context are strictly
INET definitions, defined at the system controller. They do not, for
example, bear any relationship to whether or not a given detector
assigned to the signal channel is turned on or off.
INET-HP 5890
Signal Definition
GLOBAL
SIG
1
ON
SIG
1
ON
SIG
2
OFF
ACTUAL
SETPOINT
ADDR:
8,31
ACTUAL
SETPOINT
RANGED
ACTUAL
SETPOINT
FULL
RANGE
ACTUAL
SETPOINT
89