Temperature Control

Valid setpoint ranges

Valid setpoint ranges
Table 3•1lists valid setpoint ranges for the 13 keys controlling oven and
heated zone temperatures.
Table 3-1. Valid Setpoint Ranges For Temperature Control Keys
Key
OVEN TEMP
INIT TEMP
INIT TIME
RATE
FINAL TEMP
FINAL TIME
OVEN MAX
EQUIB TIME
INJ A TEMP
INJ B TEMP
DET A TEMP
DET B TEMP
AUX TEMP
NOTE: TOTAL run time will not exceed 650.00 minutes regardless of values entere
INIT TIME
*The valid setpoint range for a Flame Ionization Detector is 0 to 450 C.
46
Valid
Setpoint Range
-80 to 450
-80 to 450
0 to 650.00
0 to 70
-80 to 450
0 to 650.00
70 to 450
0 to 200.00
0 to 400
0 to 400
0 to 400*
0 to 400*
0 to 400
,
RATE
, and
FINAL TIME
In
Increments Of
1 C
1 C
0.01 minute
0.1 /minute
1 C
0.01 minute
1 C
0.01 minute
1 C
1 C
1 C
1 C
1 C
.
Function
Oven Control
Oven Control
Oven Control
Oven Control
Oven Control
Oven Control
Oven Control
Oven Control
Zone Control
Zone Control
Zone Control
Zone Control
Zone Control