NOTE
3. To examine each sampler's IF signal, redefine which port and sampler
the network analyzer uses for the phase lock. Use the following steps:
a. Redefining to Port 2, sampler a2, press:
PARAMETER

then
{USER 3 a2}

{DRIVE}

DRIVE: {PORT 2}

{PHASE LOCK}

{a2}

{REDEFINE DONE}

Observe the resulting a2 IF trace on the display.
b. Redefining to Port 2, sampler b2, press:
then
{USER 2 b2}

{DRIVE}

DRIVE: {PORT 2}

{PHASE LOCK}

{a2}

{REDEFINE DONE}

Observe the resulting b2 IF trace on the display.
4. Connect an open (or short) to Port 1 and Port 2. Power is reflected back
to b1 and b2 samplers through the coupler.
To check each sampler trace, press:
{USER 1 a1}

{USER 2 b2}

{USER 3 a2}

{USER 4 b1}

5. Record your results for paths 1 thru 4 (
See Figure 5-10 for an RF signal-path 1 trace example. For Option 007, see
Figure 5-13.
Troubleshooting Procedures
[MENU]
{REDEFINE PARAMETER}
{REDEFINE PARAMETER}
to
{USER 1}
HP 8517B S-Parameter Test Set Manual 5-15
Troubleshooting the Test Set
) in Table 5-6.
{USER 4}