Figure 3-6. Visually Aligning Clamp And Nut Flats; Positioning The Connector; Figure 3-7. Mating The Clamp And Nut Flats - HP 8517B Operating & Service Manual

S-parameter test set
Table of Contents

Advertisement

NOTE
NOTE
The flats may actually be in any orientation, with respect to the front panel.
Figure 3-6 Visually Aligning Clamp and Nut Flats

Positioning the connector

See Figure 3-7. Maneuver the clamp over the RF connector and onto the test
port connector.
1. Hold the test cable with one hand. Use the other hand to press the clamp
gently and steadily, as you wiggle it into position straight over the RF
connector and onto the test port connector nut.
Be sure to loosen the clamp when you are slipping it over the connector.
2. Fit the internal flats in the clamp over the flats on the test port connector
nut.
3. Avoid rotating the clamp as you position it so the RF connection remains
tight (remember it loosens easily).
Figure 3-7 Mating the Clamp and Nut Flats
HP 8517B S-Parameter Test Set Manual 3-17