Service Adapter Conclusions

Table 5-5
Results Observations
If all channels look good
If one or more channels look bad
Troubleshooting the Test Set
Troubleshooting Procedures
The HP 85102 is working properly. The problem is
most likely in the source or the test set. Refer to
Procedure 4, "Unratioed Power Tests" in this chapter.
The problem is most likely in the HP 85102. Refer to
the HP 8510C On-Site Service Manual for information
about troubleshooting the HP 85102.