Hoveround MPV4 Owner's Manual page 40

Mpv4
Hide thumbs Also See for MPV4:
Table of Contents

Advertisement

Hoveround.arc
11/2/06
If a fault occurs, you can find out what has happened by counting the number of
bars on the battery gauge that are flashing.
1 bar:
2 bars:
3 bars:
4 bars:
5 bars:
6 bars:
7 bars:
8 bars:
9 bars:
10 bars:
10 bars Rapid Flash:
10 bars, Three Second Blink (Pilot) or 6 bars, three second blink (VSI):
Speed Bars Cycling (VSI) or bars cycling (Pilot):
Speed Bars Flashing:
MPV 4s with this controller system use a flashing "System Status Indicator"
light in the key symbol.
40
Chapter 9
10:07 AM
Page 40
Low battery or a bad connection.
Bad connection to left motor.
Short circuit, left motor to battery.
Bad connection to right motor.
Short circuit, right motor to battery.
Verify battery charger is not plugged in (Pilot). Controller fault (VSI).
Joystick fault.
Controller fault.
Problem with brake circuit. Check that manual brake release levers are
in the drive position and power seat lift is lowered (if equipped).
An excessive voltage to the control system. This is usually caused by a
poor battery connection.
Powered up when joystick deflected. Restart with joystick in neutral.
Control is in sleep mode. Press Off/On twice to restart.
Controller locked.
Power seat lift is inhibiting drive. Lower the seat.
System Status Indicator
Light
Figure 9A
D82008204 REV B 1/14/04

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents