Maintenance

Performance Test
Verify Probe accuracy by measuring the output of a 20 A (±0.15 %),
60 Hz current source (Fluke 5520A or equivalent). When used with a
compatible DMM with ±0.35 %, or better, ac current accuracy (Fluke
8808A or equivalent), the probe should measure 20.000 A ±0.400 A
(20.000 mA, ±0.400 mA). There are no calibration adjustments in
the Clamp.
Cleaning
Clean the case periodically by wiping it with a damp cloth and
detergent. Do not use abrasive cleaners or solvents. Do not
immerse the Clamp in liquids.

Contacting Fluke

To contact Fluke, call:
USA: 1-888-993-5853
Canada: 1-800-363-5853
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.

Specifications

Electrical Characteristics
Specified Current Range
Overload Capability for 1 Hour,
33 % Duty Cycle
Output Sensitivity
Turns Ratio
Accuracy
Load Impedance
Phase Shift at 2 kHz
Bandwidth
Working Voltage
(see Safety Standards section)
General Characteristics
Maximum Conductor Size
Output Connection
Operating Temperature Range
Storage Temperature Range
100 mA to 800 A rms
1500 A rms (50/60 Hz)
1 mA / A
1000:1
±2 % of reading, 50 Hz to
1 kHz, ±3 % of reading
(typical), 30 Hz to 50 Hz or
1 kHz to 10 kHz (Maximum
load 10 )
20
±1 degree typical
30 Hz to 10 kHz (-3 dB)
600 V ac rms or dc
54 mm diameter
4 mm safety plugs
-10 to +55 °C
-40 to +70 °C