Appendix B
Options Menu Features
Setting the Time and Date
Volume Calibration
Page 116
Programming Features
1. From the Main Menu, select
11:00 AM 21 - APR - 01
TIME / DATE
ADVANCED
OPTIONS
READY TO START
Optional Programs
•
Time and Date for internal real-time clock
•
Calibrate the sample volume
•
Advanced Features
•
Adjust level readings of optional integral flow meter
1. From the Main Menu, select
11:00 AM 21 - APR - 01
ACCEPT
CLEAR
ENTRY
USE +/- KEY TO CHANGE 12/24 HR FORMAT
2. Adjust the real time clock and date setting, using
keys.
CHANGE MONTH
From the Main Menu, select
The liquid sensor on the sampler offers unparalleled sample accuracy and
repeatability. For most site conditions, no volume calibration is necessary, and
the volume collected is accurate within 10% of the requested volume. To
improve volume accuracy to 5%, or when sample volume accuracy is
adversely affected by unusual site conditions (such as the sample source
located higher than the sampler pump), or when the liquid sensor is disabled,
then volume calibration should be performed.
Sensor Disabled
When the sensor is disabled, you must calibrate the sample volume(s). If
more than one volume is programmed (such as when using storm water or
upset sampling), you must calibrate each volume independently (see chart
below).
.
OPTIONS
OPTIONS>TIME/DATE
_ _: _ _ AM _ _ -APR- _ _
MODE: 12-HR FORMAT
OPTIONS> VOLUME CALIBRATION
OPTION MENU
VOLUME
CALIBRATION
LEVEL
ADJUST
TIME / DATE
CHANGE
AM / PM
CHANGE
MONTH
and
CHANGE AM/PM
.
8990apdx_pfeatures.fm