Hach 900 MAX Instrument Manual page 55

Refrigerated sampler
Table of Contents

Advertisement

Step 15 - Start/Stop Times
8990startup.fm
Bottle Number
If the Program Complete Output is disabled, then it is used in conjunction with
this Special Output to transmit the bottle number to the connected device.
The Special Output signal can be configured to activate during one of the
following conditions:
After Each Sample—One second pulse at the completion of each sample
cycle. It signals an external datalogger or PC that a sample cycle was
initiated. When this choice is enabled, the sample success/failure is also
transmitted to the external datalogger by means of Pin F on the
Auxiliary connector.
Only When Pumping—During sample intake portion of cycle only,
ignoring all purges and rinse cycles. Used for tipping solenoid or ball
valves when sampling from a pressurized line.
From Rinse To Purge—During entire sample cycle, including all purge
and rinse cycles. Used for tripping solenoid or ball valves when sampling
from a pressurized line.
Start/Stop Times start and stop a program at pre-arranged times. For
example, this feature can be used to run a program during weekdays and stop
the program during weekends, restarting on the following Monday. It can also
halt sampling for nighttime shutdowns.
Up to 12 Start and 12 Stop time entries can be set in any one program. Either
a Start Time and Date or a Start Time and Day of the Week may be selected.
15-A. Highlight the Start/Stop Time choices on the Advanced Sampling menu.
Press
to continue.
SELECT
11:00 AM 21 - APR - 01
PGM CMPLT OUTPUT
SETPOINT SAMPLING
SELECT
SPECIAL OUTPUT
START/STOP TIMES
STORM WATER
TIMED BOTTLE SETS
RETURN
15-B. Enable or Disable Start/Stop Times using the
Press
to continue.
ACCEPT
15-C. Press
CHANGE CHOICE
Select Time/Date if the start and stop times are longer than one week.
Select Time/Weekday if the program repeats on a daily or weekly basis.
15-D. Press
to continue.
ACCEPT
ADVANCED SAMPLING
CHANGE CHOICE
to select either Time/Date or Time/Weekday.
Section 3
key.
Page 53
Advanced Sampling

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sigma 900 max

Table of Contents

Save PDF