Note: Conductivity temperature
calibration is only necessary when
logging temperature.
Table 12 Conductivity Values at Temperature for Hach KCl Solution
Solution
Calibration Value
Temp °C
to be Entered
30
29.8
29.6
29.4
29.2
29
28.8
28.6
28.4
28.2
28
27.8
27.6
27.4
27.2
27
26.8
26.6
26.4
26.2
26
25.8
25.6
25.4
25.2
Page 78
Conductivity Probe
c. If the calibration temperature is lower than the labeled value, then
subtract that value from the standard (1.0 mS) to get the actual value
to be used for calibration.
d. If the calibration temperature is higher than the labeled value, then
add that value to the standard (1.0 mS) to get the actual value to be
used for calibration.
7. Using the value that was calculated in step 6, enter the conductivity of the
solution then press
Calibrating the Conductivity Temperature
1. Place the probe in a liquid and wait for the temperature reading
to stabilize.
2. Enter the actual temperature of the liquid (the current reading is shown for
reference). Temperature calibration is complete.
Solution
Temp °C
1.099
25
1.095
24.8
1.091
24.6
1.087
24.4
1.083
24.2
1.079
24
1.075
23.8
1.071
23.6
1.067
23.4
1.063
23.2
1.059
23
1.055
22.8
1.051
22.6
1.047
22.4
1.043
22.2
1.039
22
1.035
21.8
1.032
21.6
1.028
21.4
1.024
21.2
1.020
21
1.016
20.8
1.012
20.6
1.008
20.4
1.004
20.2
1.0 mS - (correction factor) 0.13002 = 0.86998 mS
to complete the calibration.
ACCEPT
Calibration Value
to be Entered
1.000
0.996
0.992
0.988
0.984
0.980
0.976
0.972
0.968
0.965
0.961
0.957
0.953
0.949
0.945
0.941
0.937
0.933
0.929
0.925
0.921
0.917
0.913
0.909
0.905
Solution
Calibration Value
Temp °C
to be Entered
20
0.902
19.8
0.898
19.6
0.894
19.4
0.890
19.2
0.886
19
0.882
18.8
0.878
18.6
0.874
18.4
0.870
18.2
0.866
18
0.862
17.8
0.858
17.6
0.854
17.4
0.850
17.2
0.846
17
0.842
16.8
0.838
16.6
0.835
16.4
0.831
16.2
0.827
16
0.823
15.8
0.819
15.6
0.815
15.4
0.811
15.2
0.807