Appendix B: Remote Interface Reference
CONFigure:VOLTage:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for AC voltage measurements with the specified range and
resolution. This command does not initiate the measurement. For AC measurement, resolution is
fixed at 6 1/2 digits, therefore the resolution parameter only affects the front panel display.
CONFigure:CURRent:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC current measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:CURRent:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for AC current measurements with the specified range and
resolution. This command does not initiate the measurement. For AC measurement, resolution is
fixed at 6 1/2 digits, therefore the resolution parameter only affects the front panel display.
CONFigure:RESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 2-wire Ohms measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:FRESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 4-wire Ohms measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:FREQuency {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for frequency measurements with the specified range and
resolution. This command does not initiate the measurement. For frequency measurements, the
meter uses only one "range" for all inputs between 3Hz and 300kHz. With no input signal applied,
frequency measurements return "0."
CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range and resolution. This
command does not initiate the measurement. For period measurements, the meter uses only one
"range" for all inputs between 0.33sec and 3.3µsec. With no input signal applied, period
measurements return "0."
CONFigure:CONTinuity
Preset and configure for a continuity measurement. This command does not initiate the
measurement. The range and resolution are fixed at 1k
B-8
Model 2100 6 1/2-Digit Resolution Digital Multimeter User's Manual