Section 6: Error Messages
Table 6-1 (continued)
Execution error codes
Error code
531 Insufficient memory
532 Cannot achieve
requested resolution
540 Cannot use overload as
math reference
550 Command not allowed in
local
6-4
Model 2100 6 1/2-Digit Resolution Digital Multimeter User's Manual
Description
There is not enough memory to store the requested number of readings in
internal memory using the
SAMPle:COUNt
count (
not exceed 512 readings.
The multimeter cannot achieve the requested measurement resolution.
You may have specified an invalid resolution in the
MEASure
command.
The multimeter cannot store an overload reading (9.90000000E+37) as the
math reference for null or dB measurements. The math state is turned off
as a result of this condition.
The multimeter received a
Return to
Section Topics
INITiate
command. The product of the sample
TRIGger:COUNt
) and the trigger count (
CONFigure
READ?
command while in the local mode.
2100-900-01 Rev. B / July 2007
) must
or