Refer to Figure 30:
F
S
UNCTION 
ETTING 
Press the MENU button and use the up/down arrow keys to scroll to FUNCTION
Function menu
Description
Recall Factory
Set the meter's settings to their factory default
conditions
Auto Calibration
Perform an Auto-calibration procedure
Language
Select the desired display language (English, Chinese,
Deutsch, or Greek)
Auto-calibration
The Auto-calibration function automatically configures internal parameters to maintain the
sensitivity and accuracy of the meter. Use the Auto-calibration in the following cases:
When the temperature changes by more than 5 degrees Celsius during operation
When operating the oscilloscope in a new bench top or field environment
1. Press MENU and select the FUNCTION menu using the MENU ▲ or MENU ▼ key
2. Press F2 (Auto calibration). A message appears asking the user to remove all cables
 
Figure 30: Save and Recall Waveforms
M
(
T
ENU 
REFER TO 
ABLE BELOW
51 