8. Calibration

8-1. Calibration Group

The balance has the following modes as a calibration group.
Calibration
Automatic self calibration (Calibration due to changes in temperature for GX-K series)
Calibration using the internal mass for GX-K series (One-touch calibration)
Calibration using an external weight
Calibration Test
Calibration test using an external weight (Calibration test does not perform calibration)
Correction of the internal mass value
Terms
The following terms are defined as follows:
Internal mass
External weight
Calibration weight
Target weight
Caution
The GF-K series does not perform "automatic self calibration" and "one-touch calibration
using the internal mass", as it does not include an internal mass.
Calibration adjusts the balance for accurate weighing.
Besides periodic calibration and before each use, perform calibration when:
The balance is installed for the first time.
The balance has been moved.
The ambient environment has changed.
Do not allow vibration or drafts to affect the balance during calibration.
To output the data for GLP using the RS-232C interface, set "GLP output (
"Data output (
dout
the GLP report concerning the GX-K series. If the time or date is not correct, adjust them.
Refer to "10-7 Clock and Calendar Function".
Calibration test is available only when "GLP output (
"1" or "2",
For GX-K series, the calibration and calibration test data can be stored in memory. To store
them, set "Data memory (
For GF-K series, the calibration and calibration test data is not stored in memory.
The value of the internal mass may change due to aging, corrosion or other damage
caused by the operating environment. Check the internal mass periodically. Correct the
internal mass value as necessary.
= Built-in calibration weight (GX-K series only)
= A weight that you have. Referred to as a calibration weight when
used for calibration.
= A weight used for calibration
= An external weight used for calibration test
)". Refer to "10. Function Table". The time and date can be added to
)" to "
". Refer to "12. Data Memory" for details.
data
3
)" of "Data output (
info
24
)" of
info
)" is set to
dout