Accumulation Function - A&D GP-61KS Instruction Manual

Gp series precision balance
Table of Contents

Advertisement

5-5 Accumulation Function

The accumulation function adds the weighing data and displays the total value.
To use the accumulation function, set the "Accumulation function (add)" parameter of the function
table as described below.
Note
While the accumulation function is in use, the data memory function is not available. When
using the accumulation function, make sure that the "Data memory (data)" parameter of the
function table is set to "0".
Selecting the unit
1 Press the MODE key to select a unit to be used for accumulation.
Note
While the accumulation function is in use, unit selection using the MODE key is not
available.
Setting the function table
2 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release
the key.
3 Press the PRINT key.
4 Press the SAMPLE key several times to display add 0 .
5 Press the RE-ZERO key to display add 1 .
Note
To disable the accumulation funciton, set the "Accumulation function (add)" parameter to "0".
6 Press the PRINT key to store the setting.
7 Press the CAL key to return to the weighing mode.
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents