Adding The Comparison Results; Main Display Comparison Function - A&D GP-61KS Instruction Manual

Gp series precision balance
Table of Contents

Advertisement

Cp lnt
9 With
displayed with all the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to proceed to step10.
When the current setting is to be changed, press the RE-ZERO key. The balance is now in
the digital input mode. Change the setting using the following keys.
SAMPLE key
RE-ZERO key
PRINT key
CAL key
Notes: Enter the tolerance value in percentage, with the reference value as 100%.
Only the digital input mode is available for setting the tolerance value.
The MODE key is not used to set the tolerance value.
10 Press the CAL key to exit the comparator function and return to the weighing mode.
Notes
When Pound/Ounce is selected as a weighing unit, enter the values in ounces for
comparison.
In the density mode, comparison is performed to the density obtained.

9-11 Adding the Comparison Results

By setting "Comparison results (Cp-r)" of the function table to "1", the comparison results can be
added to the data output using the RS-232C serial interface. Use A&D standard format (type 0).
The comparison results are added after the header in A&D standard format as below.

9-12 Main Display Comparison Function

The main display comparison function displays the comparison results in a magnified way, on the
main portion of the display in place of the weight value.
Selecting a unit
1 Press the MODE key to select a unit to be used for comparison.
Note
While the main display comparison function is in use, unit selection using the MODE key is not available.
displayed, press the PRINT key. The current setting of the tolerance value is
To select the digit to change the value.
To change the value of the digit selected.
To store the new setting and go to step 10.
To cancel the new setting and go to step 10.
52