Smart Range Function - A&D GP-61KS Instruction Manual

Gp series precision balance
Table of Contents

Advertisement

3-6 Smart Range Function

The GP-32K, GP-32KS and GP-102K are equipped with two ranges. The precision range has a
higher resolution. The standard range has normal resolution.
The range is switched automatically, depending on the value displayed. Pressing the RE-ZERO key
allows weighing in the precision range, regardless of the tare value. (Smart range function)
The range can be fixed to the standard range, by pressing the SAMPLE key.
Note
Once the range is switched to the standard range, it will not switch to the precision range
automatically even when the displayed value becomes within the precision range value. Use
the RE-ZERO or SAMPLE key to activate the precision range again.
Below is a description of how the Smart Range actually functions, using the GP-102K
(Precision range=61 kg×1 g, Standard range=101 kg×10 g)
1 Press the RE-ZERO key.
The balance will start weighing, using the
precision range.
2 Place a container, on the weighing pan, to be
used as a tare.
The value displayed exceeds the precision
range value. So, the range will be switched to
the standard range.
3 Press the RE-ZERO key.
The balance will be switched to the precision
range.
4 Place a sample on the pan.
The value displayed is within the precision
range value. So, the balance will perform a
weighing, using the precision range.
Precision range/standard range value
Precision range
(after RE-ZERO is pressed)
Standard range
GP-32K/GP-32KS
Up to 6.1009 kg
6.101 to 31.008 kg
12
GP-102K
Up to 61.009 kg
61.01 to 101.08 kg

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents